GasAlertMax
Troubleshooting
Troubleshooting
The detector’s electronics are protected from variations in
humidity and corrosive atmospheres. If you encounter a
problem, try the solutions listed in the following table.
If you still are unable to correct the problem, contact
BW
Technologies.
Table 15. Troubleshooting Tips
Problem Possible Cause Solution
The detector does not activate. → No battery
→ Depleted battery
→ Damaged or defective detector
→ Install battery
→
Replace battery
→
Contact BW Technologies
The detector enters alarm
immediately when it is activated.
→ Sensor needs to stabilize
→ Low battery alarm
→ Sensor alarm
→ Pump alarm
→ Used sensor: wait 60 seconds
New sensor: wait 5 minutes
→
Replace battery
→
Replace sensor
→ If the sampling hose is attached,
determine if it is obstructed. If it is not,
clean or replace the pump filter. If this
does not work, see the
Pump Alarm
section or
contact BW Technologies
53